Hippuric Acid (Urine) HPLC

Hippuric Acid (Urine) Hplc

Hippuric Acid (Urine) HPLC

Hippuric Acid (Urine) For HPLC

 

INTENDED USE

KR119 Reagent kit for the HPLC determination of Hippuric acid in urine.

 

KIT SPECIFICATION

Contents for 100 analyses

Cat no

Quantity

Reagents

KR1191

1000 mlx1

Mobile Phase

KR1192

100 mlx1

İnternal Standart

KR1193

0.5 mlx5

Urine Calibrator


ACCESSORIES

KR1195

1 adet

HPLC column for the analysis of Hippuric acid in urine.

KR1199

100 adet

Autosampler vial with teflon caps

 

CONTROLS

KR1197

0.5 ml x5

Urine Control, Level I

KR1198

0.5 ml x5

Urine Control, Level II

 

INTRODUCTION

Hippuric acid, o-, m-, p-Methylhippuric acid, Mandelic acid and Phenylglyoxylic acid are the main metabolites of the organic solvents: toluene, xylene, and styrene. These toxic compounds are widely used throughout various industries, making the determination of their concentrations relevant to occupational health in a number of different professional fields. Due to their high levels of volatility, they primarily enter the body through inhalation and are rapidly absorbed through the lungs. For occupational health monitoring of people who are exposed to these compounds at work, measurement of the metabolites in urine is more reliable than determining the concentrations of the toxins in the blood, due to their very fast metabolism rates.

For any hazardous substance or its metabolite, a BTV (Biological Tolerance Value) is defined as a threshold value for maximum absorption in biological material.

 

EQUIPMENT AND INSTRUMENTAL PARAMETERS

The analysis of Hippuric acid requires a HPLC including gradient system withHPLC pump, autosampler, autoinjector and UV detector.

The use of a solvent degasser prevents  air bubbles and ensure a stable baseline.

 

Sampler:                        Injection volume 5 μl, analytical run time 20 min..

HPLC pump:                   Flow rate 1.0 ml/min is recommended

Column temperature:      Ambient (~25 °C)

FLD detector:                  225 nm

 

COLLECTION AND STORAGE OF PATIENT SAMPLES

The sample material is 24 hours urine. Storage life at room temperature is about 1 day, refrigerated (+2 to +8 °C) up to 1 week and frozen below -20 °C up to 6 months.

 

SAMPLE PREPARATION PROCEDURE

25 µl urine sample is added to vial with teflon cap

1 ml internal standard is added.

Vortex the mix for 10s.

Centrifuge at 14 000 rpm for 10 min.

Samples prepared as above can be stored at ambient temperature for about 24 hours, at

+2 to +8 °C up to 36 h, frozen below -18 °C up to 3 months.

 

 

 

 

 

 

 